From: Jonathan Cameron <jic23@kernel.org>
To: Dan Murphy <dmurphy@ti.com>, linux-iio@vger.kernel.org
Subject: Re: [PATCH v2 2/3] iio: bindings: Add TI afe4403 heart monitor documentation
Date: Sun, 17 May 2015 13:51:56 +0100 [thread overview]
Message-ID: <55588EEC.6000901@kernel.org> (raw)
In-Reply-To: <1431374726-23383-3-git-send-email-dmurphy@ti.com>
On 11/05/15 21:05, Dan Murphy wrote:
> Add the TI afe4403 heart monitor device tree
> binding documentation. health directory
> created under iio.
>
> Signed-off-by: Dan Murphy <dmurphy@ti.com>
A few bits popped up when reviewing the driver.
> ---
> .../devicetree/bindings/iio/health/ti_afe4403.txt | 26 ++++++++++++++++++++++
> 1 file changed, 26 insertions(+)
> create mode 100644 Documentation/devicetree/bindings/iio/health/ti_afe4403.txt
>
> diff --git a/Documentation/devicetree/bindings/iio/health/ti_afe4403.txt b/Documentation/devicetree/bindings/iio/health/ti_afe4403.txt
> new file mode 100644
> index 0000000..b196b17
> --- /dev/null
> +++ b/Documentation/devicetree/bindings/iio/health/ti_afe4403.txt
> @@ -0,0 +1,26 @@
> +* Texas Instruments - AFE4403 Heart rate and Pulse Oximeter
> +
> +The device consists of a low-noise receiver channel
> +with an integrated analog-to-digital converter (ADC),
> +an LED transmit section, and diagnostics for sensor and LED fault detection.
> +
> +Required properties:
> + - compatible: Must contain "ti,afe4403".
> + - ste-gpio: GPIO for the spi control line
> + - data-ready-gpio: GPIO interrupt when the afe4403 has data
This is an interrupt as far as the device is concerned, not a gpio. hence the
binding should reflect that. Not all interrupt capable devices are even capable
of reading out their current value, let alone acting as outputs as well!
> + - led-supply: Chip supply to the device
> +
> +Optional properties:
> + - reset-gpio: GPIO used to reset the device via HW
> +
> +Example:
> +
> +&heart_rate {
> + compatible = "ti,afe4403";
> + ste-gpio = <&gpio1 29 GPIO_ACTIVE_HIGH>;
> + data-ready-gpio = <&gpio1 28 GPIO_ACTIVE_HIGH>;
> + led-supply = <&vbat>;
> +};
> +
> +Technical Datasheet:
> +http://www.ti.com/product/AFE4403/datasheet
>
next prev parent reply other threads:[~2015-05-17 12:51 UTC|newest]
Thread overview: 8+ messages / expand[flat|nested] mbox.gz Atom feed top
2015-05-11 20:05 Heart Rate monitors v2 Dan Murphy
2015-05-11 20:05 ` [PATCH v2 1/3] iio: heart_monitors: Add an event for heart rate monitors Dan Murphy
2015-05-17 12:52 ` Jonathan Cameron
2015-06-02 16:51 ` Dan Murphy
2015-05-11 20:05 ` [PATCH v2 2/3] iio: bindings: Add TI afe4403 heart monitor documentation Dan Murphy
2015-05-17 12:51 ` Jonathan Cameron [this message]
2015-05-11 20:05 ` [PATCH v2 3/3] iio: health: Add TI afe4403 heart monitor Dan Murphy
2015-05-17 12:50 ` Jonathan Cameron
Reply instructions:
You may reply publicly to this message via plain-text email
using any one of the following methods:
* Save the following mbox file, import it into your mail client,
and reply-to-all from there: mbox
Avoid top-posting and favor interleaved quoting:
https://en.wikipedia.org/wiki/Posting_style#Interleaved_style
* Reply using the --to, --cc, and --in-reply-to
switches of git-send-email(1):
git send-email \
--in-reply-to=55588EEC.6000901@kernel.org \
--to=jic23@kernel.org \
--cc=dmurphy@ti.com \
--cc=linux-iio@vger.kernel.org \
/path/to/YOUR_REPLY
https://kernel.org/pub/software/scm/git/docs/git-send-email.html
* If your mail client supports setting the In-Reply-To header
via mailto: links, try the mailto: link
Be sure your reply has a Subject: header at the top and a blank line
before the message body.
This is an external index of several public inboxes,
see mirroring instructions on how to clone and mirror
all data and code used by this external index.