From: Jonathan Cameron <jic23@kernel.org>
To: Dan Murphy <dmurphy@ti.com>, linux-iio@vger.kernel.org
Subject: Re: [PATCH v2 2/3] iio: bindings: Add TI afe4403 heart monitor documentation
Date: Sun, 17 May 2015 13:51:56 +0100 [thread overview]
Message-ID: <55588EEC.6000901@kernel.org> (raw)
In-Reply-To: <1431374726-23383-3-git-send-email-dmurphy@ti.com>
On 11/05/15 21:05, Dan Murphy wrote:
> Add the TI afe4403 heart monitor device tree
> binding documentation. health directory
> created under iio.
>
> Signed-off-by: Dan Murphy <dmurphy@ti.com>
A few bits popped up when reviewing the driver.
> ---
> .../devicetree/bindings/iio/health/ti_afe4403.txt | 26 ++++++++++++++++++++++
> 1 file changed, 26 insertions(+)
> create mode 100644 Documentation/devicetree/bindings/iio/health/ti_afe4403.txt
>
> diff --git a/Documentation/devicetree/bindings/iio/health/ti_afe4403.txt b/Documentation/devicetree/bindings/iio/health/ti_afe4403.txt
> new file mode 100644
> index 0000000..b196b17
> --- /dev/null
> +++ b/Documentation/devicetree/bindings/iio/health/ti_afe4403.txt
> @@ -0,0 +1,26 @@
> +* Texas Instruments - AFE4403 Heart rate and Pulse Oximeter
> +
> +The device consists of a low-noise receiver channel
> +with an integrated analog-to-digital converter (ADC),
> +an LED transmit section, and diagnostics for sensor and LED fault detection.
> +
> +Required properties:
> + - compatible: Must contain "ti,afe4403".
> + - ste-gpio: GPIO for the spi control line
> + - data-ready-gpio: GPIO interrupt when the afe4403 has data
This is an interrupt as far as the device is concerned, not a gpio. hence the
binding should reflect that. Not all interrupt capable devices are even capable
of reading out their current value, let alone acting as outputs as well!
> + - led-supply: Chip supply to the device
> +
> +Optional properties:
> + - reset-gpio: GPIO used to reset the device via HW
> +
> +Example:
> +
> +&heart_rate {
> + compatible = "ti,afe4403";
> + ste-gpio = <&gpio1 29 GPIO_ACTIVE_HIGH>;
> + data-ready-gpio = <&gpio1 28 GPIO_ACTIVE_HIGH>;
> + led-supply = <&vbat>;
> +};
> +
> +Technical Datasheet:
> +http://www.ti.com/product/AFE4403/datasheet
>
next prev parent reply other threads:[~2015-05-17 12:51 UTC|newest]
Thread overview: 8+ messages / expand[flat|nested] mbox.gz Atom feed top
2015-05-11 20:05 Heart Rate monitors v2 Dan Murphy
2015-05-11 20:05 ` [PATCH v2 1/3] iio: heart_monitors: Add an event for heart rate monitors Dan Murphy
2015-05-17 12:52 ` Jonathan Cameron
2015-06-02 16:51 ` Dan Murphy
2015-05-11 20:05 ` [PATCH v2 2/3] iio: bindings: Add TI afe4403 heart monitor documentation Dan Murphy
2015-05-17 12:51 ` Jonathan Cameron [this message]
2015-05-11 20:05 ` [PATCH v2 3/3] iio: health: Add TI afe4403 heart monitor Dan Murphy
2015-05-17 12:50 ` Jonathan Cameron
Reply instructions:
You may reply publicly to this message via plain-text email
using any one of the following methods:
* Save the following mbox file, import it into your mail client,
and reply-to-all from there: mbox
Avoid top-posting and favor interleaved quoting:
https://en.wikipedia.org/wiki/Posting_style#Interleaved_style
* Reply using the --to, --cc, and --in-reply-to
switches of git-send-email(1):
git send-email \
--in-reply-to=55588EEC.6000901@kernel.org \
--to=jic23@kernel.org \
--cc=dmurphy@ti.com \
--cc=linux-iio@vger.kernel.org \
/path/to/YOUR_REPLY
https://kernel.org/pub/software/scm/git/docs/git-send-email.html
* If your mail client supports setting the In-Reply-To header
via mailto: links, try the mailto: link
Be sure your reply has a Subject: header at the top and a blank line
before the message body.
This is a public inbox, see mirroring instructions
for how to clone and mirror all data and code used for this inbox;
as well as URLs for NNTP newsgroup(s).